MarkWard - 27 Sep 2017
Previous Logbook
25-10-2017
- Lots of meetings with Sam / Cryovac, good news is that the vessel should 'should' be here today, keep fingers crossed.
- We aquired quotes from kingsbury for cranage in preparation for our own shipping options
- Speciality Spanish Transport, quote was lowered and we accepted
- Southern Cranes and Access gave an insane quote, I think this was in response to bad communications between SST and cryovac and themselves which basically involved SST demanding a quote without giving them any information about the Job. So... knowing or believing they had no options, they quoted £1500 ish... which Cryovac (as much as they are not in a position to complain really) identified as crazy.
- We provided Cryovac with our quote from Kingsbury
- Cryovac contracted Kingsbury
- Got my simple diagnoistic tool working on m3daq
- I made a sym link into /home/dmatter/bin (i think, i cant check, m3daq is offline for the delivery)
- This means the program can be used from any command line location
- Invoked by rawAna.exe - if given with no options it will print the usage directions and flags
- For looking quickly at a root file and rumping the number of events/triggers point rawAna.exe at a raw root file with no arguments
- -d dumps waveforms to screen ->really fast, much better than using a CINT macro
- -c dumps CCD images
- -d -c dumps both to screen
- It can be used for offline analysis, currently it will do baseline cuts, peak time cuts, find peakheights for the charge channels and output 4 histograms. baseline sigma cut and time cut can be added using appropriate flags (see useage)
- Wrote small script for exporting fits file using the DMTPC image library for Kulveer to use
- Lab cleanup, tidy with the team, My task was mostly disconnecting and tidying away all cables and moving the DAQ computer and electronics rack.
- Shoring props are in place, clean tent has not fallen over... yet
18-10-2017
- Practiced some mesh stretching, (mostly a helper with Zack and the team)
- Lots of intersting things learnt
- Started writing a tool for doing quick diagnostics on data files from DMTPC,
- Very very fast at dumping waveforms to screen, which is initially the most useful use of it, to check what triggers look like
- Optionally can dump CCD images along side the waveforms. So we can visually pair up waveforms and Events
- Slowly updating it to perform baseline cuts given a list of files and generally have it do all of what we usually do in Macros
- Motivation... CINT IS BAD AND SLOOOOW
- Already this code appears much faster
- TPC rail mount designs -> Not got a full assembly drawing yet, though have had a few communications with machine ship, the rails are advancing and hopefully should be out soon.
- Reviewing Drawings with Zack.
- Chamber Shipping Drama... mostly Zack doing the waving of sticks at Sam, Adriana speaking to various spanish transport companies.
- I have spoken to the company wh lifted the DMTPC vessel, they are coming tomorrow to scope the area and will quote us for a UK lift.
-
11-10-2017
- Wrote a "How too and getting started page" for people to use to do basic dmtpc anaylsis etc it is located here
- Took data sample for purpose of CR-112 calibration, See Adriana's analysis
- Discovered Trigger issues with the DAQ
- Appears to ignore trigger and pulse polarity settings (at least for the charge sensitive pre-amp)
- Settings in the hptpc_neutron template (copied from one of the ones that shipped with the setup) has triggers on charge channels effectively disabled.
- Thus it appears we are triggering only when we have a signal on the fast amplifers, ONLY negative going
- Need to investigate this, possibility of hard coding or bug in the DAQ initialisation.
- Helped (Tiny bit) with the Ring Turning -> See Zacks page
- Helped (Tiny Tiny bit) with putting clean tent curtains up
- Created a 50ohm terminator for SHV connectors for grounding the anodes
- Grounded All anodes and ground meshes
- Ramped voltage - breakdown reoccurs at same voltages (630-650v)
- Cycled the gas, first pumping down to ~3x10^-6 mbar and then back to 40mbar
- Channels now ramp to 670 without sparking, sparking comes in closer to 680V now
- Cathode performance is identical, sparking occurs at 3200V
- Slowly ramped up all channels to 680V, sparking is moderate, though approximately 0.1Hz so we can get a few events between sparks
- Need to write some error handling into my code, currently segfaults if the first event has no triggers... Thus if the first event is after a spark, it sometimes contains zero triggers.
- Going to migrate to Linappserv and make a set of compiled programs to improve analysis speed
- Have 3 data samples in different configurations for analysis
04-10-2017
- Worked with Adriana on developing the analysis to accept multiple raw files
- Can identify each peak in the spectrum based upon behavour of the CSP channel.
-
-

-
- All this is within a single run, a spark appears to trigger the change between each state. Looking through later files, the behavour returns to normal and then back to bad again
- Applying peak time cut effectively removes this bad behavour from the files.
- See Adriana's logbook for final plot showing combined files with baseline and time peak time cuts (Also preliminary gain)
-
- Shakedown of Amplifiers was good, using the test input, scope traces looked identical for all channels.
- 3 instances of bad grounding on the boards which has now been fixed. Extra insulation on the inside of the cases has been added.
- Vessel backfilled with CF4 to 40mbar(30 torr), pressure reached before backfill was 8.44x10^-7 mbar
- Drift field achieved and stable 3200V (compared to 2700V previously) Getting better
can at do 100V/cm field at least.
- Anode voltage
Bottom left starts to spark at approx 0.5-1Hz when at a voltage of 650V, 50V short of where I usually try and take data to see how everything is behaving.
-
27-09-2017
- Baseline Analysis,
- Initially performed my own baseline extraction in order to learn a few things about the raw TFile structure
- Rewrote the same analysis in the space of a few hours using the dmtpc framework
- Process is to,
- Extract the baseline from every trace and push it into a histogram for each channel
- Fit gaussian/find RMT and define a cut of accepting only waveforms that have a baseline within 8sigma
- Remove whole events that do not make the cut, Replot and refit.
-
- Pulse height analysis
- dmtpc framework is mostly defined in Pulse.hh (for anyone who is interested)
- The getPeak() function extracts the peak and subtracts for baseline offset
- Further acceptance can be performed based on getPeakTime()
- Code (and a tutorial on what each part of my macro is doing what has been passed down to Adriana for further modification/development (getPeakTime() cuts which shouldn't take much development
- Currently my analysis works only one a single file though it should be realitively simple to make it work on a string of files.
- Quite slow... might concider a compiled version rather than a macro, shakes fist at CINT
-

-

- Friday - Adriana and I did a full vacuum system and DMTPC operations tutorial,
- Made modification to HV feedthrough, changing the connection type to a push on one
- Applied Kapton insulation over any bare wire, the push on connector and length of the pin.
- Proved the HV feedthrough operates up to the full capacity of the supply (8kV) in Nitrogen, no sparks or issues
- Evacuuated vessel
- Started Testing performance of the Cremat boards, Initially tests look like boards are healthy, but we have devised a full set of tests to be performed in order to check board health and operation of different Amps in each
- Should be completed in next couple of days, if all checks are good we move onto the grounding and backfill the detector for sample runs